REPLACEMENT RTDS
If it is necessary to replace RTDs, complete RTD kits including the energy meter’s plug-in connector and calibration values for
the replacements are available from the manufacturer.
It is also possible to use other manufacturer’s RTDs. The RTDs must be 1000 Ohm platinum RTDs suitable for a three-wire
connection. A connection adapter, P.N. D005-0350-300, is available to facilitate connection to the energy version.
See Figure 37.
PIN # 1
PIN # 3
PIN # 5
PIN # 2
PIN # 4
PIN # 6
PIN # 8
R TD2
R TD1
W H ITE
R E D
B L ACK
G R E E N
B R OW N
B L UE
DR AIN
W H ITE
B L ACK
R E D
G R E E N
B L UE
B R OW N
DR AIN
PIN# 3
PIN# 6
PIN# 1
PIN# 8
PIN# 4
PIN# 2
PIN# 5
Figure 37: Ultrasonic energy - RTD adapter connections
OTE:N It will be necessary to calibrate third party RTDs to the flow meter for proper operation. See the APPENDIX of this
manual for the calibration procedure.