Series 901 Enhanced Ultrasonic Flow Meter Operations & Maintenance Manual REV 4/02
Rev. 04/02 -9- D901/M PART 2 - PRE-INSTALLATION CHECKOUT After unpacking, it is recommended to save the shipping carton and
Rev. 04/02 -10- D901/M The transducers that are utilized by the D901/M contain piezo electric crystals for transmitting and
Rev. 04/02 -11- D901/M Liquid that contains greater than 10,000 PPM [1%] of 30 micron or greater suspended solids or aeratio
Rev. 04/02 -12- D901/M The mounting location and straight pipe requirements for CASE 2 liquid characteristics are the same
Rev. 04/02 -13- D901/M PART 2 - TRANSDUCER INSTALLATION Before the transducer heads are bonded to the pipe surface, an area
Rev. 04/02 -14- D901/M 2. Wrap the strap around the pipe in the area where the transducers are to be mounted. Leave the st
Rev. 04/02 -15- D901/M the same direction for proper operation. See Figure 7. NOTE: Large pipes may require two people fo
Rev. 04/02 -16- D901/M The enclosure should be mounted in an area that is convenient for servicing, calibration or for obser
Rev. 04/02 -17- D901/M ♦ Where little vibration exist. ♦ Protected from falling corrosive fluids. ♦ Within ambient tem
Rev. 04/02 -18- D901/M enclosure. Secure the transducer cable with the supplied conduit nut. 2. The terminals on the tra
Rev. 04/02 -1- D901/M This manual contains detailed operating instructions for all aspects of the D901/M instrument. The fo
Rev. 04/02 -19- D901/M If the 0-1mA output has been installed on the D901/M, the connections are available on the screw ter
Rev. 04/02 -20- D901/M Note: The D901/M flow meter system requires a full pipe of liquid before a successful startup can be
Rev. 04/02 -21- D901/M (covered in the previous sections of this manual) the D901/M can be keypad configured to provide sele
Rev. 04/02 -22- D901/M The ID key allows the entry of a pipe’s Internal Diameter. The D901/M utilizes the I.D. constant to
Rev. 04/02 -23- D901/M The FULL SCALE key allows the entry of the highest anticipated fluid velocity. The entry is made as
Rev. 04/02 -24- D901/M The VOL. PULSE key allows the entry of a totalizer exponent. This feature is useful for accommodati
Rev. 04/02 -25- D901/M When utilizing the MGD engineering UNITS, the totalizer defaults to a VOL. PULSE multiplier of 10E3.
Rev. 04/02 -26- D901/M By applying a CAL value other than 100%, the factory calibrated readings will be altered by the perc
Rev. 04/02 -27- D901/M The DAMP key allows the selection of time duration between display updates. The value selected and e
Rev. 04/02 -28- D901/M The RESET key is used for generating a system reset or to reset the accumulated (totalized) flow.
Rev. 04/02 -2- D901/M D. Place each transducer under the mounting strap, 180° apart on the pipe. Ensure that the transduce
Rev. 04/02 -29- D901/M PART 4 - TROUBLE SHOOTING CONDITION POSSIBLE CAUSE Unit does not turn “ON” when power is applied •
Rev. 04/02 -30- D901/M PART 4 - TROUBLE SHOOTING FAULT light is ON; low SIGNAL STRENGTH indication (continued) • If GND c
Rev. 04/02 -31- D901/M PART 5 - APPENDICES Appendices Relay Alarm Module Addendum Pipe Dimension Chart: Cast Ir
Addendum — Relay Alarm Module Description: Installation and Operation Manual Addendum Series D901/D90M Alarm Relay Option Origin Date: 7-2
Addendum — Relay Alarm Module Description: Installation and Operation Manual Addendum Series D901/D90M Alarm Relay Option Origin Date: 7-2
Cement Lining Std./Double ThicknessID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall3 3.96 3.46 0.25 3.40 0.28 3.34 0.31 3.28 0.34 3.22 0.37 3.1
Size(Inches)O.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchW
ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall1 1.315 1.185 0.065 1.097 0.109 1.049 1.049 0.1
NominalPipe(Inches)I.D. INCH1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 91 1.05 2.6989 4.0484 5.3978 6.7473 8.097 9.4462 10.796 12.145 13.490 14.8
NominalPipe(Inches)I.D. INCH1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9FPS TO GPM CROSS - REFERENCE (Schedule 40)18 16.88 697.52 1046.3 1395.0 1
Rev. 04/02 -3- D901/M TABLE OF CONTENTS Pages Quick-Start Operating Instructions 1-2 Introduction General 5 Ap
Fluid Sound SpeedsOriginal Date: 10/19/99Revision: noneRevision Date: noneFile: I:/dynasonics/dyna_code/tables/doppler ss conversions.xls120.0176921Do
Linseed Oil .925-.939 1770 5803.3 118Methanol (40,41) 0.79 1076 3530.2 72Methyl alcohol (40,44) 0.79 1076 3530.2 72Methylene chloride (3) 1.33 1070 35
GENERAL TERMS AND CONDITIONS OF SALES 1. PAYMENT – Terms of payment are effective from the actual date of invoice. If, in the Seller’s opinion
Limited Warranty and Disclaimer Dynasonics, div. of Racine Federated Inc. warrants to the end purchaser, for a period of one year from the date
RETURN OF EQUIPMENT/SALES INFORMATION CONTACTS AND PROCEDURES Customer Service/Application Engineer: If you have a question regarding order statu
D901/M O&M REV 04/02 8635 WASHINGTON AVENUE RACINE, WI 53406 TOLL-FREE IN NORTH AMERICA.: TEL: (800) 535-3569 FAX: (800) 732-8354 TEL: (262) 6
Rev. 04/02 -4- D901/M TABLE OF CONTENTS Pages Instrument Startup and Configuration CAL. Constant Input 26 Displa
Rev. 04/02 -5- D901/M PART 1 - INTRODUCTION The D901/M ultrasonic flow meter is designed to measure the fluid velocity of li
Rev. 04/02 -6- D901/M PART 1 - INTRODUCTION The D901/M employs modular construction and provides electrical safety for the o
Rev. 04/02 -7- D901/M PART 1 - SPECIFICATIONS Ambient Conditions Transmitter - -22 to 160°F [ - 30 to 70 °C], 0-95% relativ
Rev. 04/02 -8- D901/M Outputs 4-20 mA, 600 Ohms max., Isolated. 12 Vdc pulse, 50 mS duration, 0-0.15 Hz to 0-2.5 Hz, user