Flow
*
**
Flow
*
**
Flow
*
**
Flow
Flow
*
**
Flow
*
**
24
24
14
10
10
10
5
5
5
5
5
5
***
Upstream
Pipe
Diameters
Downstream
Pipe
Diameters
Piping Configuration
and Transducer Positioning
TABLE 2.1  PIPING CONFIGURATION AND
TRANSDUCER POSITIONING
~ Mount the transducers in an area where they will not
be inadvertently bumped or disturbed during normal
operation.
~ Avoid installations on downward  owing pipes unless
adequate downstream head pressure is present to over-
come partial  lling of or cavitation in the pipe.
The  ow meter system will provide repeatable measure-
ments on piping systems that do not meet these require-
ments, but accuracy of these readings may be in uenced to
various degrees.
STEP 2TRANSDUCER SPACING
TFXL remote mount transit time  ow meters can be used
with four di erent transducer types: DTTN, DTTH, DTTS
and DTTC. Meters that utilize the DTTN or DTTH transducer
sets consist of two separate sensors that function as both
ultrasonic transmitters and receivers. DTTS and DTTC trans-
ducers integrate both the transmitter and receiver into one
assembly that  xes the separation of the piezoelectric crys-
tals. DTTN and DTTH transducers are clamped on the outside
of a closed pipe at a speci c distance from each other.
Transducer
Mount
Mode
Pipe Material Pipe Size
Liquid
Composition
W-Mount
Plastic (all types)
2-4 in.
(50-100 mm)
Low TSS;
non-aerated
Carbon Steel
Stainless Steel
Copper
Ductile Iron
Not
recommended
Cast Iron
V-Mount
Plastic (all types)
4-12 in.
(100-300 mm)
Carbon Steel
Stainless Steel
Copper
4-30 in.
(100-750 mm)
Ductile Iron
2-12 in.
(50-300 mm)
Cast Iron
Z-Mount
Plastic (all types)
> 30 in.
(> 750 mm)
Carbon Steel
> 12 in.
(> 300 mm)
Stainless Steel
Copper
> 30 in.
(> 750 mm)
Ductile Iron
> 12 in.
(> 300 mm)
Cast Iron
TSS = Total Suspended Solids
TABLE 2.2  TRANSDUCER MOUNTING MODES
 DTTN AND DTTH
The DTTN and DTTH transducers can be mounted in:
W-Mount where the sound traverses the pipe four
times. This mounting method produces the best relative
travel time values but the weakest signal strength.
V-Mount where the sound traverses the pipe twice.
V-Mount is a compromise between travel time and
signal strength.
Z-Mount where the transducers are mounted on oppo-
site sides of the pipe and the sound crosses the pipe
once. Z-Mount will yield the best signal strength but
the smallest relative travel time.
For further details, reference Figure 2.1. The appropriate
mounting con guration is based on pipe and liquid char-
acteristics. Selection of the proper transducer mounting
method is not entirely predictable and many times is an
TOP VIEW
OF PIPE
W-Mount V-Mount Z-Mount
TOP VIEW
OF PIPE
TOP VIEW
OF PIPE
FIGURE 2.1 TRANSDUCER MOUNTING MODES 
DTTN AND DTTH