26 06-TTM-UM-00158 8/2012
SPECIFICATIONS
Liquid Types Most clean liquids or liquids containing small amounts of suspended solids
Power Requirements 11-28VDC @ 0.25A
Protection Reverse polarity, surge suppression
Velocity 0.1…40 fps (0.03…12 mps)
Inputs/Outputs 4-20mA Output (Standard Output) Totalizer Pulse
Resolution
Power
Insertion loss
Loop impedance
Isolation
12-bit for all outputs
Source
5V Max.
900 Ohms Max.
Can share ground common
with power supply isolated
from piping system
Operation
Pulse duration
Source/sink
Logic
Normal state high; pulses low
with display total increments
30mSec min.
2 mA max.
5 VDC
Turbine Frequency Output/TTL—Pulse Output (Switch selectable)
Type
Amplitude
Frequency range
Duty cycle
Non-grounded referenced AC/ground referenced square wave
500mVpp min. /5 VDC
0…1,000Hz
50% ± 10%
Display Type: 2 line x 8 character LCD
Rate: 8 maximum digits with lead zero blanking
Total: 8 maximum digits with exponential multipliers from –1…6
Units Engineering units: Feet, gallons, ft
3
, million-gal, barrels (liquid & oil), acre-feet. lbs, meters, m
3
, liters, million-liters, kg
Rate: second, minute, hour, day
Enclosure – Rating:
Dimensions:
NEMA 3 (Type 3) ABS, PVC and Ultem® (integral system), brass or SS hardware
3"W x 2.5" H x 6" D (75 mm x 63 mm x 150 mm)
Transducer Type: Clamp-on, uses time of  ight ultrasonics
Ambient
Temperature:
General purpose: –40° F…185° F (–40° C…85° C); Hazardous locations integral mount: 0° F
105° F (–20° C…40° C); Hazardous locations DTTN: –40° F…185° F (–40° C…85° C)
Transducer Ratings: DTTN/DTTC: NEMA 6* (IP 67), CPVC, Ultem®, Nylon cord grip, PVC cable jacket; -40…250° F (-40…121° C)
Construction DTTN: NEMA 6P* (IP 68) option, CPVC, Ultem®, Nylon cord grip, Polyethylene
cable jacket; -40…250° F (-40…121° C)
DTTH: NEMA 6* (IP 67), PTFE, Vespel®, Nickel-plated brass cord grip, PFA cable jacket; -40…350° F (-40…176° C)
DTTS: NEMA 6* (IP 67), PVC, Ultem®, Nylon cord grip, PVC cable jacket; -40…185° F (-40…85° C)
*NEMA 6 units: to a depth of 3 ft. (1 m) for 30 days max. NEMA 6P units: to a depth of 100 ft.
(30 m) seawater equivalent density inde nitely.
Pipe/Tubing Sizes: 1/2" (12 mm) and larger
Pipe/Tubing
Materials:
Carbon steel, stainless steel, copper and plastic
Accuracy: DTTN/DTTH ±1% of reading at rates >1 FPS (0.3 MPS), ±0.01 FPS (±0.003 MPS) at rates lower than 1 FPS;
DTTS/DTTC 1” and larger units ±1% of reading from 10-100% of measuring range, ±0.01 FPS (±0.003 MPS) at
rates lower than 10% of measuring range; ¾” and smaller units ±1% FS.
Refer to the Dimensional Speci cations page for applicable measuring ranges for each DTTS/DTTN transducer models.
Repeatability: ±0.5% of reading
Response Time: 0.3…30 seconds, adjustable
Certi cations: All TFXL Models
General Requirements: UL 61010-1 and
CSA C22.2 No. 61010-1
Hazardous Locations: Class I Div. 2 Groups C&D
T4 to UL 1604 and CSA 22.2 No. 213
DTTN Transducer and IS Barrier (-F option)
Hazardous Location Designation: Class I Div 1,
Groups C & D; T5 Intrinsically Safe Exia
Process Control Equipment: CSA C22.2 No. 142
Intrinsically Safe Equipment: CSA C22.2 No. 157
Intrinsically Safe & Associated Apparatus: UL913
Energy Management Equipment: UL916
ULTRALINK™ Utility: Software utility, requires serial communication cable
Windows 2000, Windows XP, Windows Vista®, and Windows® 7 compatible