TFXL ERROR CODES
Revised 5-25-2009
Code Number Description Correction
Warnings
0010
Signal Strength is below Signal Strength
Cuto entry
Low signal strength is typically caused by one of the
following:
Empty pipe
Improper programming/incorrect values
Improper transducer spacing
Non-homogeneous pipe wall
0011
Measured Speed of Sound in the liquid is
greater than ±10% di erent than the value
entered during meter setup
Verify that the correct liquid was selected
in the BASIC menu.
Verify that pipe size parameters are correct.
Class C Errors
1001 System tables have changed
Initiate a meter RESET by cycling power or by selecting
SYSTEM RESET in the SEC MENU.
1002 System con guration has changed
Initiate a meter RESET by cycling power or by selecting
SYSTEM RESET in the SEC MENU.
Class B Errors
3001 Invalid hardware con guration Upload corrected le.
3002 Invalid system con guration Upload corrected le.
3003 Invalid strategy le Upload corrected le.
3004 Invalid calibration data Re-calibrate the system.
3005 Invalid speed of sound calibration data Upload new data.
3006 Bad system tables Upload new table data.
Class A Errors
4001 Flash memory full Return unit to factory for evaluation
TABLE A5.1 TFXL ERROR CODES