Rev. 5/09 -1.11- TFXP
1. The connectors located on the side of the TFXP consist of three
1/4-turn BNC-type and one 5.5 mm power plug. These
connectors are environmentally sealed, but it is recommended
not to allow water or other liquids to collect in the electrical
connections pocket.
2. Connect the appropriate wires to the corresponding connections
on the transmitter. The transducer cable has markings of
UPSTREAM and DOWNSTREAM to assist in the installation
process. The UPSTREAM transducer is the one located closer
to the direction from which fluid flow normally comes from. (The
fluid normally passes the UPSTREAM transducer before passing
the DOWNSTREAM transducer.) If the transducer wires are
connected backwards, a negative flow indication will be
observed on the flow meter display. See Figure 1.5 or the wiring
diagram located on the inner door of the transmitter.
NOTE: The transducer cable carries low level, high frequency
signals. In general, it is not recommended to add additional cable to
the cable supplied with the DTTN, DTTH, DTTS or DTTC
transducers. If additional cable is required, contact the Dynasonics
factory to arrange an exchange for a transducer with the appropriate
length of cable. Cables to 990 feet (300 meters) are available. If
additional
cable and connections are added, ensure that they
are RG59 75
Ohm compatible.
PART 1 - TFXP TRANSMITTER CONNECTIONS
Electrical
Connections
It is highly recommended that the internal battery in the TFXP be
fully charged before using the meter for the first time. Details
covering this procedure are located on page 1.1 of this manual.
Figure 1.5 — Transmitter Connections