Rev. 5/09 - 2.2 - TFXP
The first step in the installation process is the selection of an
optimum location for the flow measurement to be made. For this to
be done effectively, a basic knowledge of the piping system and its
plumbing are required.
An optimum location is defined as:
A piping system that is completely full of liquid when
measurements are being taken. The pipe may become
completely empty during a process cycle – which will result in
the error code 0010 (Low Signal Strength) being displayed on
the flow meter while the pipe is empty. Error codes will clear
automatically once the pipe refills with liquid. It is not
recommended to mount the transducers in an area where the
pipe may become partially filled. Partially filled pipes will cause
erroneous and unpredictable operation of the meter.
A piping system that contains lengths of straight pipe such
as those described in Table 2.1 on page 2.3. The optimum
straight pipe diameter recommendations apply to pipes in both
horizontal and vertical orientation. The straight runs in Table 2.1
apply to liquid velocities that are nominally 7 FPS (2.2 MPS). As
liquid velocity increases above this nominal rate, the requirement
for straight pipe increases proportionally.
Mount the transducers in an area where they will not be
inadvertently bumped or disturbed during normal operation.
Avoid installations on downward flowing pipes unless
adequate downstream head pressure is present to overcome
partial filling of or cavitation in the pipe.
1. Mounting Location
PART 2 - TRANSDUCER & RTD INSTALLATION