Rev. 5/09 -3.18- TFXP
File Number
or Location
Data Logger
Configuration
NOTE: The logger will not automatically go to the next location if
the previous location is filled. In this case when the location exceeds
30,000 data points the oldest points will be discarded in favor of new
points. This is the classic FIFO memory stack.
Up to 16 separate logging sessions are possible with up to 12,500
points each but for each new measurement session the logger must
be stopped, a new page selected using the keypad or data logger
utility, and then the logging must be started again. Similarly if a
single data acquisition session were to exceed 30,000 points the
logger must be stopped, a new location selected using the keypad
or data logger utility, and then the logging must be started again.
NOTE: The data logger is not accessible using the TFX’s infrared
adapter. Communications between the data logger and computer
must be established using a directly connected RS232C or RS485
connection.
NOTE: If the data logger is accessed during a logging session the
logger will stop logging as soon as the data retrieval is started. The
logger must be started again when the retrieval process is complete.
NOTE: The flow meter must be on to supply power to the data
logger if information is to be retrieved while the logger is still
installed.
DATALOG PRESENT -- Datalog Present (Choice)
1. Yes
2. No
The TFX must be “told” the data logger is present before any data
can be collected.
DATALOG LOC -- Datalog Location (Value)
1 to 16
2. DATALOG OPERATION MENU
PART 3 - KEYPAD CONFIGURATION